Jiao, Zheng, Nibras, Ghajari, Sadeghi, Cortese, McEuen, Molnar and Lee, IEEE TBioCAS 2026 (Prof. Sunwoo Lee's group with Cornell co-authors): the FB-MOTE recording chain rebuilt as a system model, checked against the paper's measurements, improved
2-page summary (PDF) Full report (PDF, 9 pages)
Published design against the proposal
Published, measured: the paper's values (Sec. IV, Figs. 11-12, 20-21). Published, our model: the same chain in the behavioural model, with the paper's decoder of Sec. V. Proposed: the same model with the changes under By. All model numbers are simulation, read from the Python results; red rows are costs of a change.
The four changes
The proposed chain
The published chain (ink) with the four changes in the accent colour. The small plots are a browser run of this chain at 900 uW/mm2 with 10 detected photons per pulse.
The evidence in three plots
The chain, original against proposed
A recording of synthetic LFP, three spike units and biological noise through the whole FB-MOTE chain of Jiao et al. (TBioCAS 2026): the inverter-based amplifier whose bandwidth and noise follow the light-dependent current, the relaxation-oscillator PWM encoder with pseudo-chopping, the PPM pulse pairs from the micro-LED, the optical path with background light, a photon-counting detector and the decoder. Light enters only as the bias current it produces. The paper's design and the design you choose run on the same recording and the same light; every change is drawn in the accent colour on the lower schematic, and the small plots are the simulated signals of this run.
PAPER'S DESIGN
PROPOSED DESIGN
Photon budget: how few photons can the link live with?
Deeper implants and poor alignment leave fewer photons per micro-LED pulse at the detector. This sweep runs one recording through the link at each photon count and decodes it with the paper's decoder (one Otsu threshold per 100 ms batch, Sec. V) and with the proposed grid-tracking decoder (P2). Squares: the Python reference runs (2 s recordings); lines: this browser run.
The blocks, checked against the paper and in SKY130
Every system result rests on these block models. The points are the paper's measurements, digitised from its figures (about 3 % reading error); the lines are the model. The SKY130 curves are ngspice runs of transistor-level replicas (open PDK, tt corner); FB-MOTE itself is TSMC 180 nm, so the replicas check the physics, not the absolute values.
Amplifier: bandwidth against current
Amplifier: input noise density against current
SKY130 amplifier replica (Fig. 4, forward-biased bulk)
Encoder oscillator: period against current
SKY130 oscillator: sampling rate over bandwidth
LED charge pump: charge per capacitance (Eq. 13)
Qinject/Ctotal = [(N - 1) VPV - ΔV] / N2 for four values of ΔV: more stacked capacitors are not always better, which is why the paper chose N = 2 (Fig. 9).
Nine proposals, ranked from safe to ambitious
Each proposal changes one block and is tested on the same system model against the same baseline. Numbers in this table come from the Python reference runs (results files of the repository, not typed by hand); the System simulator and Photon budget tabs rerun P1 to P4 live. Negative results are kept.
| # | Proposal | Block changed | Python result | Verdict |
|---|
Validation: browser model against the Python reference
The browser model is a port of the Python model. Its random draws differ (another generator, another recording with the same statistics), so the numbers agree statistically, not digit for digit. The Python model itself was checked block by block against the paper (the first rows). This table reruns the main experiments in your browser now.
| Check | Paper | Python model |
|---|
| Experiment | Python | This browser |
|---|
How the model works
The original chain
FB-MOTE (Jiao, Zheng, Nibras, Ghajari, Sadeghi, Cortese, McEuen, Molnar and Lee, IEEE TBioCAS 2026; led by Prof. Sunwoo Lee's group at NTU with Cornell co-authors) is a 43 µm × 269 µm tetherless neural recording microsystem in 180 nm CMOS. Light reaching the chip is collected by its forward-biased bulk junctions, which supply 0.2 to 1.4 µA at 0.317 to 0.43 V. An inverter-based amplifier takes most of the extra current, so its bandwidth grows with light while the integrated input-referred noise stays near 10 µVrms (about 14 to 15 µVrms under optical powering). A relaxation-oscillator encoder turns the amplified signal into pseudo-chopped PWM, a delay and XOR make a reference pulse at each rising edge and an encoding pulse at each falling edge, and a charge pump fires an AlGaAs micro-LED for 0.05 % of each period. Outside, a detector and a decoder recover the pulse delays (Sec. V).
Block models
- Amplifier: in subthreshold, gm and Go follow the amplifier current I - Io, so the corner is kL(I - Io) and the noise density c/√(I - Io); one joint fit to Fig. 11 gives both, and their product predicts a flat integrated noise. White and 1/f noise, a 2 Hz high-pass and a first-order low-pass at the corner. Under light the noise density is 1.43 times the electrical one at the same bandwidth at every level (Figs. 11 and 12 read together).
- Encoder: period T = A I-q fitted to Fig. 11 (+50 % current gives +15 % clock, as the paper states), 1e-4 cycle jitter, mid-period sampling, pseudo-chopping, the encoder time gain of Fig. 11 (top right), a 2 % offset.
- Link: Poisson photons per pulse over 20 ns with 1 ns jitter, uniform background, photon counting in 10 ns bins.
- Paper's decoder: Sec. V re-implemented (tanh-compressed histogram, Otsu threshold, non-maximum suppression with τs = 200 ns and τp = 1 µs, shorth period, folded-onset reference seeding), with the grid re-anchored on each found pulse, the most favourable reading.
- Proposed decoder: acquisition from exact pulse triplets (robust to a changing period), a phase-locked loop on the reference pulses (second order when references are sparse), each photon cluster scored by its photon count plus a continuity prior, and imputation when nothing beats "pulse lost".
What is assumed
The photon-counting channel and its background rate, the 50 Hz flicker corner, the ±1.5 mV full scale, the 1e-4 oscillator jitter, the light-flicker profile, and the synthetic neural signal (LFP 150 µVrms, spike units of -60, -100 and -150 µV, 6.3 µVrms biological noise taken from the paper's in-vivo minus open-air noise). SKY130 is not the paper's process; replicas check the physics only. The photovoltaic conversion itself is not modelled.
Source
Paper: R. Jiao et al., "A 43 µm × 269 µm light-adaptive optoelectronic autonomous microsystem for neural recording," IEEE Trans. Biomed. Circuits Syst., vol. 20, no. 2, pp. 180-193, Apr. 2026, doi:10.1109/TBCAS.2026.3652195. Its figures and numbers belong to its authors. Everything on this page is a behavioural simulation by Nithin Purushothama; nothing was fabricated or measured.