BioZ Lab

system model of the HEEPidermis bio-impedance chain, built from block models, original against proposed reportpapernithinpuru.com

Sapriza et al., HEEPidermis, LASCAS 2026: the bio-impedance chain rebuilt as a system model from SKY130 block models, checked against the paper, then improved

The paper's design against the proposal

Paper: the published chain in this model (SKY130 VCO, iDAC and temperature data; the paper's firmware). Proposed: the same model with the change under By. All numbers are simulation, read from the Python results; nothing here is measured.

The proposed circuit

One ring VCO is multiplexed between the tissue node and the reference node every 1 ms; the reference current DAC is servoed until both phases give the same count, so conductance is the ratio of two currents and the VCO's curve and drift cancel. Changes in the accent colour. The small plots are a 30-minute run of this circuit with the chip warming onto skin.

The evidence in three plots