Sapriza et al., HEEPidermis, LASCAS 2026: the bio-impedance chain rebuilt as a system model from SKY130 block models, checked against the paper, then improved
2-page summary (PDF) Earlier detailed report (PDF)
The paper's design against the proposal
Paper: the published chain in this model (SKY130 VCO, iDAC and temperature data; the paper's firmware). Proposed: the same model with the change under By. All numbers are simulation, read from the Python results; nothing here is measured.
The proposed circuit
One ring VCO is multiplexed between the tissue node and the reference node every 1 ms; the reference current DAC is servoed until both phases give the same count, so conductance is the ratio of two currents and the VCO's curve and drift cancel. Changes in the accent colour. The small plots are a 30-minute run of this circuit with the chip warming onto skin.
The evidence in three plots
The chain, original against proposed
A 30-minute skin-conductance (GSR) recording through the whole HEEPidermis chain of Sapriza et al. (LASCAS 2026): the LDO, the electrode-skin conductance, the 8-bit current DAC, the 31-stage ring VCO-ADC with its non-reset counter, the timer and the firmware that turns counts into conductance. Each block is a model with transistor-level data behind it (SKY130 runs, see Blocks). The paper's design runs in the left column of figures; the design you choose runs on the same recording, the same chip and the same temperature. Choose the proposed design below and the lower schematic redraws, with every change in the accent colour. The small plots on the schematics are the simulated signals on those nets over the 30-minute run.
PAPER'S DESIGN
PROPOSED DESIGN
Impedance spectroscopy
The iDAC plays a sine around a DC current into the tissue (a Cole model) and into Rref = 50 kOhm; each node is read by its own VCO-ADC, sampled at 20 kHz, linearised by that channel's look-up table and demodulated at the excitation frequency. Choose the tissue and the method; the run uses two Monte Carlo channels (VCO gain error and iDAC bits).
The blocks and their transistor-level data
Every system result above rests on these block models. The curves are SKY130 ngspice runs (open PDK, per-device Pelgrom mismatch); HEEPidermis itself is TSMC 65 nm, so the SKY130 shapes are mapped onto the paper's ranges (408 to 800 mV, 36 to 887 kHz, 40 nA LSB).
VCO-ADC: f(V), 31-stage current-starved ring
The ring's frequency is exponential at the bottom and saturates at the top, which is why the paper keeps the input in a narrow window. Dashed: the same VCO at 17, 37 and 47 C (SKY130 temperature runs); dotted: the linear CCO of P7.
iDAC: 8-bit binary current sink, 40 Monte Carlo chips
DNL and INL of one chip from its eight SPICE bit currents (every unit device mismatched). These errors are what sank the multisine (P3) and the nominal-current servo (P4).
Tissue: Cole model
|Z| and phase of the default tissue (R0 100 kOhm, Rinf 20 kOhm, fc 300 Hz, alpha 0.8; assumed values for a dry-skin electrode pair).
Firmware: self-calibrated LUT (P1)
Sixteen 50 ms points through Rref = 60 kOhm at start-up give the channel's own f(V); the firmware inverts it with a monotone cubic. The line is the design curve the paper's linear firmware assumes.
Thirteen proposals, three rounds
Each proposal changes one or more blocks and is tested on the same system model against the same baseline. Numbers in this table come from the Python reference runs (results files of the repository, not typed by hand); the GSR system and Spectroscopy tabs rerun the main ones live. Negative results are kept.
| # | Proposal | Blocks changed | Python result | Verdict |
|---|
Validation: browser model against the Python reference
The browser model is a port of the Python model. Its random draws differ (another generator, another GSR trace with the same statistics), so the numbers agree statistically, not digit for digit. This table reruns the main experiments in your browser now.
| Experiment | Python | This browser |
|---|
How the model works
The original chain
HEEPidermis (Sapriza et al., EPFL ESL, LASCAS 2026) is a RISC-V SoC for bio-impedance with two measurement channels. An 8-bit current DAC (40 nA LSB, up to 200 kHz update) sinks a programmed current through the tissue from a 0.8 V LDO; a second iDAC does the same through a reference resistor. Each node drives a 31-stage ring VCO whose non-reset counter is read by a timer, so the ADC is a frequency counter: f(V) between 36 and 887 kHz for 408 to 800 mV. A level-crossing unit wakes the CPU only when the reading leaves a window. The paper's results are post-layout simulation of a taped-out 65 nm design.
Block models
- VCO-ADC: SKY130 ring f(V) shape (PCHIP on log f) mapped onto the paper's range; white-FM phase noise (0.2 % per cycle, an assumption); non-reset counter with the fine phase of the 31 taps (1/32 count); temperature from SKY130 runs at 17 to 47 C.
- iDAC: eight SPICE bit currents per chip (40 Monte Carlo chips) and the output-voltage dependence of the full scale.
- Tissue: Cole model; GSR as a slowly varying conductance with skin-conductance responses every 40 s (synthetic, bi-exponential).
- Firmware: the paper's linear conversion and the proposals' LUT, background checks, lock-in, channel swapping, null balancing and Cole fitting, run exactly as written in the Python reference.
What is assumed
Phase-noise level, tissue values, the GSR recording, the VCO power model (proportional to frequency, 20 uW at the top), the 2 % VCO gain spread between channels, 0.1 % accuracy of measured iDAC bit weights. SKY130 is not the paper's process; shapes are mapped, absolute values are not claimed.
Source
Paper: J. Sapriza et al., "HEEPidermis: a versatile SoC for BioZ recording," Proc. IEEE LASCAS 2026, doi:10.1109/LASCAS67804.2026.11457136. Its figures and numbers belong to its authors. Everything on this page is a behavioural simulation by Nithin Purushothama; nothing was fabricated or measured.